Skip to main content Skip to navigation

4 mm HX High Temp H160530

Probe description PH MASWVT850W^ BL4 N-P/H (H160530)
Temperature range: -120 oC +300 oC (note that a different set-up is required for low and high temperature experiments)

Tuning range:

1H-range X-range Comments
1H 850 MHz 31P 265-345 MHz short circuit screws unscrewed range coil mounted
1H 850 MHz 13C 152-266 MHz short circuit screw unscrewed
1H 850 MHz 15N 86-154 MHz

short circuit screw 1H turned in


Probe Test Results (Stated 90o pulse lengths are the best achieved during installation. To reduce risk of damage to the probes, facility users should use slightly lower RF nutation frequencies (i.e. longer 90o pulse lengths) where possible. Please discuss with the facility manager appropriate RF nutation frequencies for your experiments.)

Nucleus 90o Pulse (us) Power (dB) Power (W) Configuration Sample
1H 2.5 - 4 970 λ/2 Adamantane
31P 2.5 - 1.3 521 λ/2, range coil ADP
79Br 3 -0.77 460 λ/2 KBr
15N 5.5 - 3.12 792 λ/4 Glycine


Temperature test:
Probe tested at indicating temperature of 300 oC (real temperature 284 oC if calculated with a 0.025 ppm / oC slope for 79Br on KBr).

Temperature calibration performed by Kenneth Inglis (Liverpool)