4 mm HX High Temp H160530
Probe description PH MASWVT850W^ BL4 N-P/H (H160530)
Temperature range: -120 oC +300 oC (note that a different set-up is required for low and high temperature experiments)
Tuning range:
1H-range | X-range | Comments | ||
1H | 850 MHz | 31P | 265-345 MHz | short circuit screws unscrewed range coil mounted |
1H | 850 MHz | 13C | 152-266 MHz | short circuit screw unscrewed |
1H | 850 MHz | 15N | 86-154 MHz |
short circuit screw 1H turned in |
Probe Test Results (Stated 90o pulse lengths are the best achieved during installation. To reduce risk of damage to the probes, facility users should use slightly lower RF nutation frequencies (i.e. longer 90o pulse lengths) where possible. Please discuss with the facility manager appropriate RF nutation frequencies for your experiments.)
Nucleus | 90o Pulse (us) | Power (dB) | Power (W) | Configuration | Sample |
1H | 2.5 | - 4 | 970 | λ/2 | Adamantane |
31P | 2.5 | - 1.3 | 521 | λ/2, range coil | ADP |
79Br | 3 | -0.77 | 460 | λ/2 | KBr |
15N | 5.5 | - 3.12 | 792 | λ/4 | Glycine |
Temperature test:
Probe tested at indicating temperature of 300 oC (real temperature 284 oC if calculated with a 0.025 ppm / oC slope for 79Br on KBr).
Temperature calibration performed by Kenneth Inglis (Liverpool)