2.5 mm HX (+19F) H13889
Probe description: PH MASDVT850W6 BL2.5 14N-31P/F-H
Additional components: range coil (for 31P) (4 turns)
Probe Test Results (Stated 90o pulse lengths are the best achieved during installation. To reduce risk of damage to the probes, facility users should use slightly lower RF nutation frequencies (i.e. longer 90o pulse lengths) were possible. Please discuss with the facility manager appropriate RF nutation frequencies for your experiments.):
Nuc | 90opulse |
Pulse power |
Pulse power (W) |
Configuration | Sample | Remarks |
79Br | 2.0 | 2.2 | 176 | λ/2 | KBr | |
13C | 2 | 2 | 192 | λ/2 | Adaman. | line width 1.6Hz |
1H | 1.65 | -1 | 225 | λ/2 | Adaman. | there is a visible 1H background signal |
13C | 2.26 | 3.6 | 133 | λ/2 | Glycine (CP) | S/N 69 (4 scans) |
1H | 1.7 | 0 | 113 | λ/2 | Glycine | 147kHz for 50ms decoupling |
15N | 4 | 2.1 | 216 | λ/4 | Glycine (CP) (enriched) | |
15N | 3.28 | 0 | 351 | λ/4 | Glycine (CP90) (enriched) | |
14N | 2.75 | -1 | 568 | λ/4 | NH4Cl | |
19F | 1.8 | 0.4 | 171 | λ/2 | pvdf | |
31P | 1.8 | 3 | 210 | λ/2+range coil | ADP |
Tuning range 1H/X mode (measured on bench)
λ/2 mode |
213 →344 MHz Over 250MHz a range coil must be inserted |
Includes: 13C(213.733), 27Al (221.483), 123Te(222.443), 51V(223.575), 23Na(224.841), 63Cu(225.382), 81Br(229.555), 129Xe(236.387), 65Cu(241.431), 71Ga(259.222), 141Pr(260.270), 125Te(268.173), 11B(272.714), 115Sn(278.109), 87Rb(278.124), 117Sn(302.874), 119Sn(316.970), 7Li(330.342), 31P(344.086) |
λ/4 mode |
61 →168 MHz |
176Lu(69.114), 37Cl (69.324), 131Xe(70.073), 169Tm(70.465), 61Ni (75.956), 91Zr(79.019), 85Rb( 82.067), 35Cl (83.282), 135Ba( 84.443), 50V( 84.748), 15N( 86.163), 10B(91.321), 153Eu( 93.330), 137Ba( 94.460), 175Lu( 96.934), 181Ta( 101.912), 123Sb( 110.153), 133Cs(111.487), 138La( 112.152), 17O(115.230), 9Be( 119.440), 139La(120.068), 6 Li(125.087), 2H(130.480), 209Bi(136.589), 75As(145.542), 171Yb(148.744), 199Hg(152.242), 77Se(162.108), 29Si(168.871) |
Tuning range 19F/X mode (measured on bench)
λ/2 mode |
213→344 MHz Over 250MHz a range coil must be inserted |
Includes: 13C(213.733), 27Al (221.483), 123Te(222.443), 51V(223.575), 23Na(224.841), 63Cu(225.382), 81Br(229.555), 129Xe(236.387), 65Cu(241.431), 71Ga(259.222), 141Pr(260.270), 125Te(268.173), 11B(272.714), 115Sn(278.109), 87Rb(278.124), 117Sn(302.874), 119Sn(316.970), 7Li(330.342), 31P(344.086) |