7mm low gamma H13895
Probe description: PH MASDVT850W6 103Rh-17O
Probe Test Results (Stated 90o pulse lengths are the best achieved during installation. To reduce risk of damage to the probes, facility users should use slightly lower RF nutation frequencies (i.e. longer 90o pulse lengths) were possible. Please discuss with the facility manager appropriate RF nutation frequencies for your experiments.)
Nuc | 90o pulse |
Pulse power |
Pulse power (W) |
Configuration | Sample | Remarks |
14N | 6 | -1 | 500 | short circuit | NH4Cl | |
35Cl | 4.5 | -3 | 760 | 10pF C | NH4Cl | |
39K | 7.2 | -4 | 797 | short circuit | KBr |
Tuning range (on bench)
low frequency range |
21→ 93 MHz |
Includes: 179Hf( 21.394), 41K(21.771), 167Er(24.480), 155Gd(26.095), 103Rh(27.085), 57Fe(27.521), 145Nd(28.560), 161Dy(29.240), 149Sm(29.240), 73Ge(29.651), 83Kr(32.705), 177Hf(34.059), 157Gd(34.255), 107Ag(34.406), 183W(35.414), 147Sm(35.445), 87Sr (36.837), 105Pd(38.897), 99Ru(39.144), 109Ag(39.555), 39K(39.664), 163Dy(40.970), 173Yb(40.978), 89Y(41.652), 101Ru(43.872), 143Nd(46.325), 47Ti (47.919), 49Ti(47.932), 53Cr(48.046), 40K(49.317), 25Mg(52.034), 67Zn(53.183), 95Mo(55.394), 201Hg(56.198), 97Mo(56.556), 43Ca(57.205), 14N(61.424), 33S(65.246), 189Os(66.006), 21Ne(67.102), 176Lu(69.114), 37Cl (69.324), 131Xe(70.073), 169Tm(70.465), 61Ni (75.956), 91Zr(79.019), 85Rb( 82.067), 35Cl (83.282), 135Ba( 84.443), 50V( 84.748), 15N( 86.163), 10B(91.321) |
high frequency range (with 10pF serial C) |
73 → 117 MHz |
61Ni(75.956), 91Zr(79.019), 85Rb( 82.067), 35Cl (83.282), 135Ba( 84.443), 50V( 84.748), 15N( 86.163), 10B(91.321), 153Eu(93.330), 137Ba( 94.460), 175Lu( 96.934), 181Ta(101.912), 123Sb( 110.153), 133Cs(111.487), 138La(112.152), 17O(115.230) |