Skip to main content Skip to navigation

7mm low gamma H13895

Probe description: PH MASDVT850W6 103Rh-17O

Operating instructions

Probe Test Results (Stated 90o pulse lengths are the best achieved during installation. To reduce risk of damage to the probes, facility users should use slightly lower RF nutation frequencies (i.e. longer 90o pulse lengths) were possible. Please discuss with the facility manager appropriate RF nutation frequencies for your experiments.)

Nuc 90o pulse

Pulse power
level (db)

Pulse
power (W)
Configuration Sample Remarks
14N 6 -1 500 short circuit NH4Cl  
35Cl 4.5 -3 760 10pF C NH4Cl  
39K 7.2 -4 797 short circuit KBr  

 

Tuning range (on bench)

low frequency range

21→ 93 MHz

Includes: 179Hf( 21.394), 41K(21.771), 167Er(24.480), 155Gd(26.095), 103Rh(27.085), 57Fe(27.521), 145Nd(28.560), 161Dy(29.240), 149Sm(29.240), 73Ge(29.651), 83Kr(32.705), 177Hf(34.059), 157Gd(34.255), 107Ag(34.406), 183W(35.414), 147Sm(35.445), 87Sr (36.837), 105Pd(38.897), 99Ru(39.144), 109Ag(39.555), 39K(39.664), 163Dy(40.970), 173Yb(40.978), 89Y(41.652), 101Ru(43.872), 143Nd(46.325), 47Ti (47.919), 49Ti(47.932), 53Cr(48.046), 40K(49.317), 25Mg(52.034), 67Zn(53.183), 95Mo(55.394), 201Hg(56.198), 97Mo(56.556), 43Ca(57.205), 14N(61.424), 33S(65.246), 189Os(66.006), 21Ne(67.102), 176Lu(69.114), 37Cl (69.324), 131Xe(70.073), 169Tm(70.465), 61Ni (75.956), 91Zr(79.019), 85Rb( 82.067), 35Cl (83.282), 135Ba( 84.443), 50V( 84.748), 15N( 86.163), 10B(91.321)

high frequency range (with 10pF serial C)

73 → 117 MHz

61Ni(75.956), 91Zr(79.019), 85Rb( 82.067), 35Cl (83.282), 135Ba( 84.443), 50V( 84.748), 15N( 86.163), 10B(91.321), 153Eu(93.330), 137Ba( 94.460), 175Lu( 96.934), 181Ta(101.912), 123Sb( 110.153), 133Cs(111.487), 138La(112.152), 17O(115.230)