3.2 mm HXY LTMAS
Probe description: PH C.MASDVT850W6 BL3.2 C/N/H
The 3.2 mm LTMAS probe can be operated in double or triple mode with a range of different nuclei combinations. Please note that the "C/N/H" indication on the probe label is incorrect. Users should refer to the table below for a list of the standard achievable tuning ranges. Additional nuclei combinations may be made available on request, please enquire with the Facility Manager.
| Typical X-Nucleus |
Typical Y-Nucleus |
Tuning range X (MHz) |
Tuning range Y (MHz) |
X-Y insert | Jumper for Cs | Shunt capacitor | Range coil |
| 13C | 15N | 172 - 215 | 83.9 - 87.5 | 13C/15N | No | 27 pF |
No |
| 13C | 2H | 172 - 215 | 106 - 139 | 13C/15N | No | No | No |
| 13C | 17O | 172 - 215 | 106 - 139 | 13C/15N | No | No | No |
| 31P |
329 -346 |
double mode adapter | No | No | 3 turns | ||
| 258 - 251 |
double mode adapter |
No | No | 5 turns | |||
| 27Al | 218 - 245 |
double mode adapter |
No | No | 9 turns | ||
| 13C | 205 - 235 |
double mode adapter |
No | No | 10 turns | ||
| 29Si | 158 - 209 |
double mode adapter |
No | No | No | ||
| 2H | 128.5 - 148 | HY | No | 6.8 pF | n.a. | ||
| 17O | 113 - 116.5 | HY | No | 27 pF | n.a. | ||
| 15N | 77 - 93 | HY | Yes | 12 pF | n.a. | ||
| 47/49Ti | TBC | 13C/15N | Yes | 33 pF | No |