Skip to main content Skip to navigation

Static H13971

Probe description: HPBBHT 850MHz Sol5 103Rh-31P

Inserts and coils

NEW: Two ports in situ setup for static NMR/MRI probes and cell cable (10 m) (supplier ePROBE GmbH Germany).

Probe Test Results (Stated 90o pulse lengths are the best achieved during installation. To reduce risk of damage to the probes, facility users should use slightly lower RF nutation frequencies (i.e. longer 90o pulse lengths) were possible. Please discuss with the facility manager appropriate RF nutation frequencies for your experiments.)

Nuc 90o pulse

Pulse power
level (db)

Pulse
power (W)
Insert Coil Sample Remarks
13C 2.7 -2.4 530 HZ3188; 1.5pF 6.2R HZ07402; 4.5wnd; Φ 5 Adam  
13C 1.88 -2.4 530 HZ03348; 1.5pF HZ07402; 4.5wnd; Φ 5 Adam  
2H 3.25 -3.36 530 HZ03344; 40pF HZ07402; 4.5wnd; Φ 5 dPMMA  
2H 4.88 -3.36 530 HZ03348; 1.5pF H7688; 9.5wnd; Φ 5 dPMMA  
2H 3.15 -3.36 530 HZ3188; 1.5pF 6.2R H7688; 9.5wnd; Φ 5 dPMMA  
35Cl 3.08 0 339 HZ1930; 60pF H7688; 9.5wnd; Φ 5 NH4Cl  
35Cl 3.2 0 339 HZ03344; 40pF H7688; 9.5wnd; Φ 5 NH4Cl  
14N 6 -1.45 500 HZ1930; 60pF H7688; 9.5wnd; Φ 5 NH4Cl  
39K 9.5 0 321 HZ03344; 40pF H7712; 20.5 wnd; Φ 5 KBr  
39K 9.7 0 321 HZ1930; 60pF H7712; 20.5 wnd; Φ 5 KBr  
39K 17 1 255 HZ1930; 60pF 13.5wnd; Φ 10 KBr  
39K 19.75 -3 641 HZ2808; 40pF 1.1R 13.5wnd; Φ 10 KBr  
31P 1.3 -1.79 400 none; parallel matching inductance soldered 3.5wnd; Φ 5 ADP  
39K 13.38 0 321 HZ1930; 60pF 17.5 wnd; Φ 7.5 KBr  


For measurements at variable temperature the upper short shielding tube has a bell-shaped dewar for thermal insulation. This dewar gives a strong 11B signal and therefore it must be removed when measuring 11B.