12T ESI FTICR mass spectrometer

July 14, 2009.


The lab is currently building a 12 Tesla FTICR mass spectrometrer with the following features.

1. Electrospray ion source.

2. Focusing quadrupole (Q0) and mass filtering quadrupole (Q1)

3. Hexapole collision cell with linac extraction electrodes.

4. Currently a transfer hexapole (H1) is incorporated, but this will be changed to a turning quadrupole in the near future. See below.

5. A thin gate valve.

6. A long transfer hexapole. This may be changed to a sectioned hexapole for decreased time-of-flight lag effects.

7. A shimmed, capacitively coupled open cylindrical ICR cell.


12T ESI FTICR mass spectrometer


In the near future, several modifications are very likely.

1. The transfer hexapole (H1) will be replaced with a couple of hexapoles and a turning quadrupole.

2. The ESI source will be mounted at 90 degrees from the current arrangement.

3. One or more other chambers will be added. In the case below, a MS/MS chamber for MAF, PD, or ETD is included, but it could also be a MALDI source if one is deemed necessary.


12T ESI FTICR mass spectrometer