Skip to main content Skip to navigation

Methods reporting

Microscope (Make, Model) Nikon CSU-W1 spinning disk with SoRa upgrade
Inverted or upright Inverted
Objective: manufacturer, magnification, numerical aperture, immersion, correction (if any)

Select the one/s used:

Nikon, 100x, 1.49, oil, CFI SR HP Apo TIRF

Nikon, 63x, 1.40, oil, CFI Plan Apo

Nikon, 40x, 1.30, oil, CFI Plan Fluor

Nikon, 40x, 0.6, air, S Plan Fluor ELWD DIC N1

Nikon, 20x, 0.5, air, CFI Plan Fluor

Light source(s)

Select the one/s used:

405 nm, 488 nm, 561 nm, 638 nm lasers

Filters, dichroic mirrors, beam splitters

Select the one/s used:

Dichroic: 405/488/561/640 nm

Secondary dichroic (when using dual-camera): 561 nm LP

Emission filters:

Blue - 446/60

Green - 525/50

Red - 600/52

FRed - 708/75

Information on other filters are available on request.

fpbase.org channel set-up https://www.fpbase.org/microscope/ZMBg8xKHJmynSD33m8RuFj/#
Additional hardware used, e.g., stage motor, incubation chamber and set-up, Piezo stage or objective

Yokogawa CSU-W1 spinning disk unit with 50 um and SoRa disks

Nikon Perfect Focus autofocus

Okolab microscope incubator

Nikon motorised xy stage

Nikon 200 um z-piezo

Bruker MiniScanner photomanipulation unit

Acquisition software and settings NIS-Elements
Camera manufacturer, model 2 x Photometrics 95B Prime