Quality control

Lateral diffraction limitAxial diffraction limitJan 2018Apr 2018Jul 2018Oct 2018Jan 2019Apr 2019Jul 2019Oct 2019Jan 202000.20.40.60.81
Figure 1. Optical performance monitoring by measuring the width of sub-resolution fluorescent beads. The Rayleigh limit is defined here as the distance between the central peak and the first minimum. Measured for n = 5 beads in x (red), y (blue) and z (green) axes.
Jan 2018Apr 2018Jul 2018Oct 2018Jan 2019Apr 2019Jul 2019Oct 201951015202530
405488561640

Figure 2. Monthly measurement of the fluorescence LED illumination power output at the microscope back aperture. "Wavelength (nm)" is the wavelength at which power is measured.